72  Willy Schauwers

              Per control weight a form for daily    EQUIP-01–01 Daily control balance’. Indicate
            weight control is available (see ‘FORM-SOP-  for each balance which control weight should be
            LABORATORY-EQUIP-01–01-Daily control     used (see ‘FORM-SOP-LABORATORY-EQUIP-
            balance’).                               01–04-Balances and control weights’). Keep all
              On this form, the following control weight   information in the logbook.
            specifications are mentioned:              If the accepted limit values are exceeded, take
                                                     the following steps.
            •  mass of the control weight and mass mini-
              mum/maximum limits, as listed by the   •  Check if the balance is completely level and
              manufacturer                             if all weight measurements are properly per-
            •  make and type.                          formed.
                                                     •  If there is still a problem, perform an internal
            cErtIFIEd controL WEIGHt                   calibration of the balance (not all balances do
            The certified control weights should be checked   have this feature).
            on a 3-yearly basis by a certified laboratory (a   •  If still a problem, contact the supervisor and
            certificate should be provided). The certified   check the balance with a certified control
            control weights are usually kept by the quality   weight.
            management department or by the head of the   •  If still a problem, contact the manufacturer.
            laboratory.                              •  If the limit values are still exceeded, the possible
                                                       consequences on earlier weight measurements
            HoW to PErForM tHE quaLIty aSSurancE       should be considered. All measurements after
            There should be quality assurance at three lev-  the last successful control weight measure-
            els:                                       ment and before the exceeded measurement
                                                       should be traced. Possible consequences on
            •  daily control by the laboratory personnel by   those laboratory results should be examined
              measuring the control weights            and rectified.
            •  half yearly control by the laboratory person-  •  The investigation  of  the  problem and all
              nel by measuring the control weight and the   corrective actions are registered on ‘FORM-
              certified control weight of every balance  SOP-LABORATORY-EQUIP-01–02-Corrective
            •  if possible, a yearly servicing by the manufac-  actions’.
              turer of the balance.
                                                     HaLF yEarLy controL
            daILy controL                            The balance is controlled half yearly by measur-
            Every day (when the balance is used), before   ing the control weight as well as the certified
            the first use in the morning, the control weight   control weight. Both weights are measured five
            should be measured and recorded on the daily   times. The average difference between both
            control sheet. Do not touch the control weight   weights has to be within a certain limit (see
            with bare hands, always use gloves or forceps   ‘FORM-SOP-LABORATORY-EQUIP-01–05-
            to handle it. Per balance and per control weight   Half yearly control of control weights’). This
            accepted weight limit values are set (accord-  limit value is determined by each laboratory
            ing to the manufacturer of the control weight   depending on the use of the balance. Balances
            or on the basis of weight results gathered over   used for preparing standard solutions have to
            a long period and the standard deviation on   be more accurate than balances used for weigh-
            these results), see ‘FORM-SOP-LABORATORY-  ing powders for bacteriology media preparation.