Gravity Metal Detector System—Operator Setup Manual


         5.2  Faults
































            Fault                   Possible Cause                                 Corrective Action


         24V           Power supply not functioning properly     Power supply needs to be checked and/ or
                                                                     replaced.


         Air Pressure    The  air  pressure  sensor  connected  to  the    Verify air pressure going to the sensor with the
                         isolated input of the DSP board (PL10 pins   reject device set to factory default; typically about
                         9 & 10) is indicating a low pressure which   60‐80 P.S.I.
                         may impair the proper function of the re‐
                         ject device.                             Verify air pressure sensors are installed.
                                                                  Verify air pressure proper functionality.
                       Air pressure sensor not functioning  prop‐
                         erly .                                   Faults can be cleared and masked in the software,
                                                                     until a technician has verified and corrected the
                                                                     causes of the fault

         Balance       There is metal located within the Detector    Verify that no metal is located within the Detector
                         aperture.                                   aperture.

                       The Detector balance is excessive.       Check if LD3 (yellow LED) on DSP board is on.

         Date Time     Indicates the time is not set.           Set the correct date and time.


         Excess        The  number  of  rejects  has  exceeded  the   Check rejected product for metal contaminants.
         Reject          limit  set  within  the  software  Feature  is     Make sure the Detector is calibrated for the
                         enabled via Fault Setup Menu.
                                                                     running product
                                                                  Can be cleared via the Clear soft key (S3)