Internal standard method

    To overcome the variation in chromatographic conditions (i.e.
       flow rate, pressure, etc.) between runs

    ● The internal standard must

      1- be similar in structure to the sample component

      2- not overlap with the sample component peak

    ● The internal standard solution is added to each standard
       solution (to construct the calibration graph) and to the sample

    ● A calibration curve is prepared by plotting
      the standard/internal standard peak area ratio (Y-axis) versus
      the concentration of the standard (X-axis).