                                                                            DUOKOPT

                                                                            Usability Testing






                     2.3.2. DIFFICULTIES MET

              Wrong position of the fingers
              Success criteria   The activation of the pumping mechanism is considered successful when the
                                 participant  put  his/her  fingers  in  accordance  with  the  instructions  or  if
                                 he/she places his/her fingers without touching the bottle tip.
              Failure criteria   When the method of activating the pumping mechanism may contaminate
                                 the bottle.
              Description        The  participant  puts  the  bottle  upside  down  and  activates  the  pumping
                                 mechanism. However, he/she didn’t follow the instructions and puts his/her
                                 fingers on the bottle tip to activate pumping mechanism.
              Explanation        One  participant  chose  to  use  the  device  with  a  hand  position  that  he
                                 considered as more adapted to him.
              Position of the    The bottle tip between the index and the middle finger, the thumb presses
              fingers            on the bottle.
              Number of          1
              participants

                     2.3.3. DISCUSSION AND CONCLUSION

              The participants expressed a feeling of natural position of the fingers and the facility of pushing on
              the device. The question 4 showed:

                  1.  The instructions stated in the leaflet are clear and well‐respected.
                  2.  The position of the fingers and the handling are easy.

                  3.  All participants easily pressed down on the device.
                  4.  Thanks to the tabs, participants handled the bottle without touching the bottle tip.


              In conclusion, the shape of the device helps the patients to activate the device correctly without

              any major difficulties.