 6.6.5  Completion of the measurements
           #          Problem            Possible cause          Solution
 •  Before unplugging the measurement cable from the device   Try pressing the ”ON/ OFF”
 receptacle disconnect the lip clip and the connection hook from the   The button is   button several times. If the
 patient.                              functioning       device still does not turn
 •  Move the file stopper to the selected reference point on the tooth.  The device does not turn   incorrectly.  on, please contact your
 •  Gently remove the file from the canal and measure the apical length   2  on by pressing the “ON /   distributor.
 between the stopper and the file.  OFF” button.  The battery is
                                       discharged.       Charge the battery.
 Note
 As a safety precaution in order to avoid over-instrumentation, it is   Electronic   Contact your distributor.
 recommended to proceed as follows: stop file progression in the   malfunction.
 canal at the point where the Propex Pixi™ indicates ‘0.0’.   3  The device shuts off during   The battery is low.  Charge the battery.
 Place the file onto an endodontic ruler and measure the apical length.   the procedure.
 Subtract a minimum of 0.5 mm from the measured file length.   No sound during the   The sound control   Adjust the sound level by
           4                           is set at “Mute”   pressing the “VOLUME”
               procedure.              level.            button.
 6.7  Sound Adjustment
                                       There is not a good   Ensure a good contact
 Propex Pixi™ is equipped with an audio indicator which enables   contact between the   between the mucosa and
 monitoring of the progression of the file within the canal in addition to   lip clip and the oral   lip clip (Place the lip clip in
 visual monitoring.                    mucosa.           the labial angle opposite
                                                         the tooth to be treated).
 The volume can be adjusted to four different levels: mute, low, normal   The connection   Clean the connection hook
 and high, by successively pressing the    “VOLUME” button.   hook is soiled.  (with Ethanol).
 When sound level is muted the    icon is turned off. At other sound
 levels the    icon remains turned on.  Deep caries
               Display is not steady   provides a        Block the external
           5                           conductive path   conductive path.
               during the procedure.
                                       outside the canal.
                                                         Remove the file, close the
                                                         perforation and repeat the
                                       Perforation.      apex detection procedure,
                                                         carefully inserting the file
                                                         into canal.
                                                         Try continuing the
                                       Large lateral canal.  procedure by gently
                                                         advancing the file.











 20/222  B ML PXEM DFU PRT / Rev.11 / 02-2018 (Old ZF 1902118.X)  B ML PXEM DFU PRT / Rev.11 / 02-2018 (Old ZF 1902118.X)  33/222  #                                                   25268-EYAL - 25268_pnim_eyal | 1 - B | 18-05-15 | 13:58:38 | SR:-- | Cyan   25268-EYAL - 25268_pnim_eyal | 1 - B | 18-05-15 | 13:58:38 | SR:-- | Mage