Page 32 - fif pnim
P. 32

#

        „   Specifications:                                                                     1.  Indications for use
        Dimensions:           120 × 74 × 70 mm                                                 CanalPro Apex Locator is an electronic device used for apex localization and working length determin-
                                                                                               ation during root canal treatment. The device enables to obtain correct results in canals with different
        Weight:               300 gr.
                                                                                               conditions (dry, wet, blood etc.).
        Type of screen:       Color Graphic TFT
 25566_eyal-Y-MK1-550 - 25566_pnim_Y-MK1-550.1 | 1 - B | 18-06-11 | 09:04:00 | SR:-- | Magenta
        Screen dimensions:    3.5”
                                                                                                2.  Contraindications
 25566_eyal-Y-MK1-550 - 25566_pnim_Y-MK1-550.1 | 1 - B | 18-06-11 | 09:04:00 | SR:-- | Yellow
 #25566_eyal-Y-MK1-550 - 25566_pnim_Y-MK1-550.1 | 1 - B | 18-06-11 | 09:04:00 | SR:-- | Black
 25566_eyal-Y-MK1-550 - 25566_pnim_Y-MK1-550.1 | 1 - B | 18-06-11 | 09:04:00 | SR:-- | Cyan
        Supply:               2.4V NiMH rechargeable battery
                                                                                               The CanalPro Apex Locator is not recommended for use in patients that have a pacemaker or other im-
        External charger:     Input: 100-240 V / 50-60 Hz                                      planted electrical devices.
                              Output: 6 V DC
                                                                                                3.  Warnings
         14.  Standard symbols                                                                 •  This product must only be used in hospital environments, clinics or dental offices by qualified dental   english
        On the device label appear standard symbols as follows:                                  personnel.
                                                                                               •  Use only the original charger.

               Class II equipment
               Type BF applied part                                                             4.  Precautions

               Direct current                                                                  Do not use CanalPro Apex Locator near devices emitting electromagnetic noise such as fluorescent
                                                                                               lamps, film viewers, ultrasonic devices, etc. Cellular phones, remote controls or other devices generating
               Manufacturer                                                                    electromagnetic waves may cause abnormal operation of CanalPro Apex Locator. Such devices should be
               Consult instructions for use                                                    turned off.
               Recycling: Please do not throw away! This product and all its components must be    During device operation protect CanalPro Apex Locator from occasional liquid spillage.

               recycled through your supplier                                                  Do not use CanalPro Apex Locator in the presence of flammable materials.
                                                                                               The CanalPro Apex Locator device should be used with the manufacturer‘s original accessories only.
                                                                                               In order to prevent infectious agent transmission it is highly recommended to use a rubber dam system
                                                                                               during the endodontic procedure.
                                                                                               To ensure that short circuits do not impair the measurements, be particularly careful with patients fitted
                                                                                               with metal crowns or bridges (avoid any metallic contact with the file or the lip clip).
                                                                                               •  High concentrations of sodium hypochlorite may result in a lower accuracy of the measurements. For
                                                                                                 working length determination, we recommend to use sodium hypochlorite solution at maximum 3 %
                                                                                                 concentration.
                                                                                               •  Make sure that the canal is wet enough to ensure reliability of the measurement.
                                                                                               •  Ensure that the file does not touch another instrument.
                                                                                               •  Avoid excessive liquids inside the tooth cavity to prevent overflow and incorrect measurements.
                                                                                               •  CanalPro Apex Locator must be stored at normal temperature (< 60 °C) and humidity conditions.
                                                                                               •  Teeth with open apices will give imprecise results.
                                                                                               •  An apex locator may not be able to measure in all conditions. In any case, it is recommended to take an
                                                                                                 X-Ray prior to use of the device and to compare the results obtained by both methods.
                                                                                               •  For your own safety, please use personal protection means (gloves, mask).


        32                                                                                                                                                          21
   27   28   29   30   31   32   33   34   35   36   37