Page 49 - Basic Monitoring in Canine and Feline Emergency Patients
P. 49

Changing conditions or personnel may also cre-  cmH O by multiplying it by a conversion factor of
                                                             2
             ate inconsistency in blood pressure monitoring. It is   1.36. However, this approach to blood pressure
  VetBooks.ir  important  to  be as  consistent  as  possible  when   measurement is not advisable as it is difficult to
             measuring serial blood pressures.  The patient’s
                                                         accurately measure the distance between the cuff
             position (e.g. left lateral recumbency), limb utilized
                                                         tioning will likely be difficult to recreate for subse-
             for the measurement (e.g. right forelimb), and cuff   and the right atrium. Furthermore, precise posi-
             size should be accurately recorded. In addition, the   quent monitoring leading to poor repeatability of
             environment, patient’s disposition or attitude, and   the measurement between time points.
             indirect monitoring device should all be docu-  All oscillometric measurements must be inter-
             mented.  The  staff  obtaining  the  measurements  is   preted carefully.  The unit provides systolic and
             also important as positioning of the patient, cuff   diastolic pressures as well as MAP. When evaluating
             location, and cuff tightness may vary between per-  the results, the clinician must consider the quality of
             sonnel. Subsequent checks should take pains to   the  pulse  pressure  on  physical  examination.  The
             recreate a similar environment and condition each   palpated pulse may be interpreted as strong, fair,
             time a blood pressure is measured to reduce   poor, or absent, dependent on the situation. When
             variability.                                the oscillometric unit displays the systolic and dias-
               Frequent mistakes include the choice of cuff size,   tolic pressures, a calculated pulse pressure can be
             patient positioning, and the type of indirect moni-  determined from the measurements (see Equation
             toring equipment used. Cuff size is important as   2.2). On occasion the machine may provide either
             using too large a cuff will generally result in falsely   very wide/large or very narrow/small pulse pres-
             low blood pressure readings. Conversely, using a   sures, and this data should be compared carefully
             cuff that is too small will result in falsely increased   with the clinician’s impression of the pulse pressure
             blood  pressure  measurements.  Cuffs  that  are  too   noted during examination.  As mentioned previ-
             small transmit pressure inefficiently to the underly-  ously, a normal or strong pulse pressure is approxi-
             ing tissues, meaning a greater pressure than appro-  mately 40-60 mmHg. If the subjective assessment of
             priate is necessary to occlude the arterial blood   the pulse strength is inconsistent with the calculated
             flow. The opposite is the case when too large a cuff   pulse pressure, the result displayed on the oscillo-
             is selected for the patient.                metric unit should be interpreted cautiously.
               Blood pressure measurement (i.e. the location of   Likely attributable to difficulty sensing and assess-
             the cuff) should occur at the level of the right atrium   ing the arterial wall oscillations, cats and small dogs
             which is generally estimated by the location of the   (<10 kg) should not have their blood pressure meas-
             thoracic inlet in lateral recumbency or the point of   ured with oscillometric equipment as this tool tends
             the shoulder in sternal recumbency. Patients having   to underestimate blood pressure. There is also some
             a blood pressure measured while  in a sternal or   controversy as to the interpretation of Doppler
             standing position often have significant vertical dis-  blood  pressure  readings in  small  dogs  and  cats.
             tance between the cuff site (usually on a limb) and   Some sources state that Doppler blood pressure
             the right atrium. This leads to false elevation of the   readings in these animals (especially in hypotensive
             blood pressure due to the weight of the blood col-  patients) may be more representative of MAP than
             umn between the right atrium and the cuff site. If a   systolic blood pressure. However, there is no consist-
             blood column exists between the cuff site and the   ent research data to support this claim.
             right atrium, the measured blood pressure includes
             both the actual blood pressure (generated by left
             ventricular contraction) and the weight of the verti-  Patient factors
             cal blood column which exerts force on the cuff.  Patients with peripheral limb disease affecting the
               An adjustment for the cuff distance (in centime-  vasculature or interstitium may have unreliable
             ters; cm) below the right atrium may be incorpo-  readings by either indirect method.  Any state of
             rated to adjust the blood pressure and appropriately   shock may create pulses with a lower pulse pres-
             decrease the measured blood pressure.  The cuff   sure, leading to a greater degree of difficulty in
             distance below the right atrium (measured in cm)   pulse  detection  and  measurement.  Furthermore,
             can be subtracted from the measured blood pres-  states of obstructive shock (such as gastric dilata-
             sure (in centimeters of water; cmH O). The meas-  tion and volvulus) can compromise blood flow to
                                         2
             ured blood pressure (in mmHg) can be converted to   the caudal aspect of the body and make blood


             Blood Pressure Monitoring                                                        41
   44   45   46   47   48   49   50   51   52   53   54