Page 849 - Small Animal Internal Medicine, 6th Edition
P. 849

CHAPTER 49   Disorders of the Endocrine Pancreas   821


            well. However, different blood glucose results during the day   400
            versus the night should be suspected if polyuria and poly-
  VetBooks.ir  dipsia persist despite blood glucose concentrations that are   300
            close to acceptable during the day, especially if polyuria and
            polydipsia are worse at night. For these cases, obtaining a
            24-hour blood glucose curve or use of a continuous glucose   Serum glucose (mg/dL)  200
            monitoring device should be considered.
              Blood glucose concentrations are typically determined by   100
            either a point-of-care glucose analyzer or handheld PBGM
            device. The accuracy of commercially available PBGM
            devices designed for use in human diabetics varies consider-  0
            ably when used in diabetic dogs, compared with results using   8 AM  12 PM  4 PM  8 PM  Mid  4 AM  8 AM
            standard reference methods (i.e., glucose oxidase and hexo-
            kinase methods) (Cohn et al., 2000; Cohen et al., 2009; Wess   FIG 49.5
            and Reusch., 2000). Blood glucose values determined by   Blood glucose concentration curves in a fractious Terrier-
            most PBGM devices designed for use in human diabetics are   mix. The same dose of NPH insulin was given for each
            typically lower than actual glucose values determined by   curve. One glucose curve (blue line) was obtained with the
            reference methods and the difference between the actual   dog in an agitated state, requiring physical restraint each
                                                                 time a blood specimen was obtained; blood for the other
            glucose value and value obtained from the PBGM increases   glucose curve (red line) was obtained through a jugular
            as hyperglycemia worsens. This bias may result in an incor-  catheter with minimal to no restraint and with the dog in a
            rect  diagnosis  of  hypoglycemia  or  the  misperception  that   quiet state. ↑, Insulin administration and food.
            glycemic control is better than it actually is. Failure to con-
            sider this rror could result in insulin underdosage and the
            potential for persistence of clinical signs despite apparently   used by clients. The biggest problem has been overzealous
            acceptable blood glucose results. One exception is the Alpha-  owners who monitor blood glucose concentrations too fre-
            TRAK® by Abbott Laboratories, a PBGM device designed for   quently and begin to interpret results and adjust the insulin
            use in diabetic dogs and cats. Accuracy of this PBGM device   dose  without  consulting  their  veterinarian,  a  practice  that
            is very good, but glucose values may be higher or lower than   ultimately leads to insulin overdosage. The reader is referred
            glucose values measured by reference methodologies on the   to page 835 for more information on home blood glucose
            same blood sample, forcing the veterinarian to accept the   monitoring.
            blood glucose concentration at face value (Cohen et al.,
            2009). Haematocrit may also affect the results of PBGMs. In   Continuous Glucose-Monitoring Systems
            one study, results of the AlphaTRAK® were less accurate   Continuous glucose-monitoring (CGM) systems are fre-
            compared with a laboratory reference method in blood   quently used to monitor glucose concentrations in diabetic
            samples with a lower haematocrit (<30%) but not an   humans and are beginning to be used in diabetic dogs and
            increased haematocrit (>50%), whereas results from a PBGM   cats. CGM systems measure interstitial fluid glucose con-
            for use in humans was less accurate with increased haema-  centrations rather than blood glucose concentrations. The
            tocrit but not a decreased haematocrit (Paul et al., 2011).  correlation between interstitial and blood glucose concen-
                                                                 trations is good. A frequently used CGM system (Guard-
            Home Blood Glucose Monitoring                        ian REAL-time, Medtronic, Northridge, Calif) measures
            Hyperglycemia induced by stress, aggression, or excitement   interstitial glucose with a small, flexible sensor inserted
            is the single biggest problem affecting accuracy of the serial   through the skin into the subcutaneous space and secured
            blood glucose curves (Fig. 49.5). The biggest factors causing   to the skin. Interstitial glucose is detected via the glucose
            stress-induced hyperglycemia are frequent hospitalizations   oxidase reaction, and detection occurs entirely at the
            and multiple venipunctures. An alternative to hospital-  electrode within the sensor component. Glucose results
            generated blood glucose curves is to have the client generate   are transmitted by a wireless transmitter to a pager-sized
            the blood glucose curve at home using the ear or metacarpal,   monitor. The interstitial fluid glucose concentration is
            metatarsal or foot pad prick technique and a PBGM device   recorded and stored every 5 minutes, and the data can be
            that allows the client to touch the drop of blood on the ear   downloaded to a computer for analysis (Fig. 49.6). Calibra-
            or foot pad with the end of the glucose test strip. There are   tion of the CGM system is required at initiation and once
            several excellent online sites that demonstrate home blood   every 12 hours. The working glucose range for the CGM
            glucose monitoring techniques for the owner of a diabetic   system is 40 to 400 mg/dL. Studies to date suggest that
            dog (e.g., Abbott Laboratories website for the AlphaTRAK2®).   the primary advantage of continuous glucose monitoring
            This technique should be considered for diabetic dogs in   is detection of hypoglycemic periods that are not detected
            which the reliability of blood glucose results generated in the   with serial blood glucose curves and a PBGM device. (See
            veterinary hospital is questionable. Home blood glucose   Suggested Readings for more information on continuous
            monitoring is also becoming a routine monitoring technique   glucose monitoring.)
   844   845   846   847   848   849   850   851   852   853   854