Traction Angle Settings









             • J, I settings – commonly used for axial distraction during the acute phase of care.

                                  •     *NOTE – The fulcrum assembly is not generally used with these traction arm settings. The fulcrum
                                        strap may be removed by sliding each end of the strap from the fulcrum strap posts.

                                  •     **NOTE – Traction pressure in the iTrac software will generally be 15-25 lbs for these traction arm
                                        settings, depending on patient stature (large patients = 20-25lbs, small/average patients = 15-20lbs).