Page 116 - Med Plaza and Cancer Center
P. 116
"Oil-Less” Medical Air Medical Systems
1.0 General Information (continued)
1.3 Tankmount Standard Components
Compressor Package
®
The LifeLine Medical Air system is fully compliant with NFPA 99. The "oil-less" duplexed compressors are
continuous duty rated with permanently lubricated, sealed bearings. The design is single stage, air-cooled,
reciprocating type with corrosion resistant reed type valves with stainless steel reeds. Both the compression
rings and rider rings are made from a long life, fluororesin material designed for continuous duty operation.
The crankshaft is constructed of a durable nodular graphite cast iron and is fully supported on both ends by
heavy duty ball bearings permanently lubricated and sealed. The crankcase is constructed of gray cast iron, not
aluminum. Maximum heat dissipation is achieved through cast aluminum alloy cylinders treated for optimum
corrosion and wear resistance. Cylinder sleeves are not required. Additionally, an insulated "heat cut" piston
pin minimizes heat transmission from the piston wall to the piston pin needle bearing. The connecting rod is of
a one-piece design for maximum reliability.
Compressor Drive
The compressors are v-belt driven through a combination compressor flywheel/sheave and steel motor sheave
with tapered bushing protected by an OSHA approved, totally enclosed belt guard. The maximum compressor
rotating speed does not exceed 1150 RPM. Provisions for v-belt adjustment are provided.
Motor
The motor is a NEMA high efficiency type, open dripproof, 1800 RPM, with 1.15 service factor suitable for
208/230/460V electrical service.
Control System
The duplex control system is NEMA 12 and U.L. Labeled. This control system provides automatic lead/lag
sequencing with circuit breaker disconnects for each compressor with external operators, full voltage motor
starters with overload protection, redundant 120V control circuit transformers, visual and audible reserve unit
alarm with isolated contacts for remote alarm, hand-off-auto lighted selector switches, automatic alternation of
both compressors with provisions for simultaneous operation if required, automatic activation of reserve if
required, visual alarm indication for high discharge air temperature shutdown with isolated contacts for remote
alarm, and duplexed runtime hourmeters.
Air Receiver
The air receiver is integrally mounted horizontally, ASME code stamped, National Board certified, and rated
for a minimum 150 PSIG design pressure. The air receiver is equipped with a liquid level gauge glass, safety
valve, pressure gauge, and automatic timed solenoid drain valve. The duplexed compressor/motor subbases are
fully isolated from the air receiver and from the steel support structure by means of a four-point heavy-duty
spring isolation system for a minimum of 95% isolation efficiency.
Intake Piping
Each compressor has a piped intake with one "hospital type" inlet air filter with threaded opening for remote
intake connection.
1-5