"Oil-Less” Medical Air Medical Systems
6.0 Inspection/Replacement Procedures
6.2 Air Intake Filter
WARNING:
BEFORE STARTING ANY MAINTENANCE PROCEDURES, DISCONNECT ALL POWER
TO THE PACKAGE.
Release all pressure from the package before removing, loosening, or servicing any covers, guards,
fittings, connections, or other devices.
Never perform any maintenance functions while the unit is in operation.
The air intake filter element should be changed every 4,000 hours of operation under normal operating
conditions. To change the filter (Ref. Fig 6-4):
1. Turn off the compressor being serviced and lock open the appropriate disconnect switches.
2. Close intake isolation valve
3. Remove the protective cover by loosening the wing nut (if applicable) and latches.
4. Remove the element.
5. Clean inside of housing.
6. Insert a new element (note orientation of the element).
7. Replace protective cover and tighten wing nut (if applicable) and latches.
8. Open intake isolation valve
9. Turn on the compressor
Canister Housing
Protective Cover
Element
Fig. 6-4 Air Intake Filter
6-6