"Oil-Less” Medical Air                                                              Medical Systems


               6.0 Inspection/Replacement Procedures


               6.2 Air Intake Filter

                                                           WARNING:
                    BEFORE STARTING ANY MAINTENANCE PROCEDURES, DISCONNECT ALL POWER
                                                      TO THE PACKAGE.

               Release all pressure from the package before removing, loosening, or servicing any covers, guards,
               fittings, connections, or other devices.

               Never perform any maintenance functions while the unit is in operation.

               The  air  intake  filter  element  should  be  changed every 4,000 hours of operation under normal operating
               conditions. To change the filter (Ref. Fig 6-4):

               1.   Turn off the compressor being serviced and lock open the appropriate disconnect switches.
               2.   Close intake isolation valve
               3.   Remove the protective cover by loosening the wing nut (if applicable) and latches.
               4.   Remove the element.
               5.   Clean inside of housing.
               6.   Insert a new element (note orientation of the element).
               7.   Replace protective cover and tighten wing nut (if applicable) and latches.
               8.   Open intake isolation valve
               9.   Turn on the compressor

                                Canister Housing

                     Protective Cover



















                                         Element

               Fig. 6-4 Air Intake Filter











                                                             6-6