Page 208 - Med Plaza and Cancer Center
P. 208

"Oil-Less” Medical Air                                                              Medical Systems


               Appendix A - Desiccant Dryer


               A.4 Maintenance

               A.4.1 Maintenance Schedule









































                          1-5 hp Duplex (Type 1 Dryer) 10-25 cfm
                Service Interval   Part number       Description      Quantity
                  Every year      ELM 22 001          Pre-filter         2
                  Every year      ELM 22 001         After-filter        2
                  Every year      MUF 01 006        Purge muffler        2
                 Every 5 years    DES 02 001        Desiccant Kit        2
                 Every 5 years    ORG 01 007        Poppet o-ring        4
                 Every 5 years    GKT 10 005        Tower gasket         4
                 Every 5 years    GKT 10 004       Canister gasket       4
                 Every 5 years    GKT 10 003       Manifold o-ring       4
                 Every 5 years    GKT 10 002     Orifice plate gasket    4
                   As needed      VAL 11 001       Switching valve       2
                   As needed      GAG 01 032    Tower pressure gauge     2
                           Purge valve (Refer to diagram for correct valve)
                 Every 5 years    VAL 04 018         Diaphragm           2
                 Every 5 years    VAL 04 029           Poppet            2





                                                            A-9
   203   204   205   206   207   208   209   210   211   212   213