Page 210 - Med Plaza and Cancer Center
P. 210
"Oil-Less” Medical Air Medical Systems
Appendix A - Desiccant Dryer
A.4.1 Maintenance Schedule
15 hp Duplex, 7-1/2 hp Triplex (Type 2 Dryer) 55 cfm
Service Interval Part number Description Quantity
Every year ELM 22 002 Pre-filter 2
Every year ELM 22 002 After-filter 2
Every year MUF 01 006 Purge muffler 2
Every 5 years DES 01 001 Desiccant 4 bags
Every 5 years VAL 01 014 Check valve 4
Every 5 years ORG 01 009 Tower o-ring 4
Every 5 years ORG 01 010 Canister o-ring 4
As needed VAL 11 001 Switching valve 2
As needed GAG 01 032 Tower pressure gauge 2
Purge valve (Refer to diagram for correct valve)
Every 5 years VAL 04 018 Diaphragm 2
Every 5 years VAL 04 029 Poppet 2
20 hp Duplex, 10 hp Triplex, 7-1/2 hp Quad (Type 2 Dryer) 75 cfm
Service Interval Part number Description Quantity
Every year ELM 22 003 Pre-filter 2
Every year ELM 22 003 After-filter 2
Every year MUF 01 007 Purge muffler 2
Every 5 years DES 01 001 Desiccant 4 bags
Every 5 years VAL 01 014 Check valve 4
Every 5 years ORG 01 009 Tower o-ring 4
Every 5 years ORG 01 010 Canister o-ring 4
As needed VAL 11 002 Switching valve 2
As needed GAG 01 032 Tower pressure gauge 2
Purge valve (Refer to diagram for correct valve)
Every 5 years VAL 04 018 Diaphragm 2
Every 5 years VAL 04 030 Poppet 2
15 hp Triplex, 10 hp Quad (Type 2 Dryer) 105 cfm
Service Interval Part number Description Quantity
Every year ELM 22 003 Pre-filter 2
Every year ELM 22 003 After-filter 2
Every year MUF 01 007 Purge muffler 2
Every 5 years DES 01 001 Desiccant 6 bags
Every 5 years VAL 01 014 Check valve 4
Every 5 years ORG 01 009 Tower o-ring 4
Every 5 years ORG 01 010 Canister o-ring 4
As needed VAL 11 002 Switching valve 2
As needed GAG 01 032 Tower pressure gauge 2
Purge valve (Refer to diagram for correct valve)
Every 5 years VAL 04 018 Diaphragm 2
Every 5 years VAL 04 030 Poppet 2