Page 282 - Med Plaza and Cancer Center
P. 282
Medical Lubricated Vacuum
Appendix A: TotalAlert Embedded Control System
A.2.10 5.7” History Screen
Figure A.13 Settings Screen
The settings screen (Figure A.13) allows the
selection of various sub screens that pertain to
system confi guration data. All value/adjustment
Figure A.12 History Screen changes are password protected.
The history screen (Figure A.12) shows all of • Vacuum – Allows the adjustment of system
the system event history excluding service vacuum operating range.
maintenance history.
• Audible – Allows the horn to re-initiate if an
• Most recent events are shown on the fi rst alarm or shutdown event has not cleared, but
(default) page and are shown in descending the horn was silenced. The allowable values
date/time order. Located at the top right are for re-initiate time are never, 15 min, 30 min, 1
the page number and the total number of hour, 8 hours, and 24 hours.
pages . The maximum number of pages is 200
(1000 events divided by 5 events per page). • Language – Allows the choice of pre-defi ned
display language.
• Page up and Page down arrows change the
view 5 events up or down at a time and are • Units – Allows the display units to be displayed
only visible if subsequent event information is and changed.
available. The Home button changes the view
to the top or most recent events and the End • Model – Displays the pertinent system model
button changes to the bottom or the oldest information.
events.
• Date/Time – Allows the date/time to be
displayed and changed.
• A maximum of 1000 most recent events are
stored for the system.
• Network – Allows the network (IP address, etc)
to be displayed and changed.
A.2.11 5.7” Settings Screen
• General – Allows the miscellaneous general
information to be displayed and changed.
A-7 4107 9000 96.03