Medical Lubricated Vacuum
Appendix A: TotalAlert Embedded Control System
A.3.9 3.5” Shutdown Screen A.4 Password Access
Figure A.23 Password Screen
The TotalAlert Embedded control system contains
Figure A.22 3.5” Shutdown Screen items that can be changed or adjusted. Before
making any changes on the system or unit screens,
The shutdown screen (Figure A.22) shows all of the a password prompt appears. A proper password
unit shutdown event information. A shutdown is must be entered before adjusting the control
classifi ed as an event of signifi cance that shuts the system. See Figure A.23.
unit down. These shutdown events are latched
and are not cleared until a user presses the reset The password for the Lubricated Rotary Vane
button on the shutdown screen. This reset button Medical Vacuum System is “121212”. After
will reset all shutdown events for that given unit. entering the password, the password remains
valid for approximately ten minutes or until the
CAUTION: The unit will need to be placed back unit resets itself to the default display screens.
in Automatic for the pump to resume normal Reentering the password after “timing out” allows
operation. the user to continue making system adjustments.
• Green shutdown condition indicates a normal A.4.1 Main Screen Password Items
status for that condition.
On the 5.7” main screen, several screens allow
• Red alarm shutdown indicates an abnormal adjustments with the proper password.
status for that condition.
• Maintenance - Resetting a Service Activity: A
service notifi cation activates when a service
item is past due. Upon completion of the
past due service activity, reset the service due
counter by pressing the item on the service
screen, selecting reset, and entering the
password when prompted.
4107 9000 96.03 A-12