Page 291 - Med Plaza and Cancer Center
P. 291

Medical Lubricated Vacuum


           Appendix A:  TotalAlert Embedded Control System



           A.7.3   Set Up: Network Confi guration






















           (Network screen is found in Settings section from
              Main toolbar)
           Figure A.27 System IP Address

           1.  IP Address using  DHCP method:  Upon
              power-up of the system, the device will
              search for a DHCP server.  If a DHCP server is
              found, IP Address, subnet mask and gateway
              are automatically obtained.  If not found, the
              device will check for a DHCP server every 30
              seconds.  When found, the IP Address, subnet
              mask and gateway are automatically obtained.

           2.  IP Address using Static method:  Upon power-
              up of the system, the device will immediately
              begin using the fi xed IP confi guration.


           A.7.4   Set Up: Connecting to the Embedded
                  Website of the Vacuum System

           1.  Start a web browser such as Microsoft Internet
              Explorer.

           2.  Enter the system’s IP address in the browser’s
              address bar.

           Example: http://10.213.11.97

           Note: To learn the IP address of the vacuum
           system, go to the 5.7” main touchscreen, press
           the Settings icon, press Network from the listing
           and the IP address will appear in the second line
           underneath the method. See Figure A.27.

           4107 9000 96.03                                 A-16