Medical Lubricated Vacuum


            Appendix A:  TotalAlert Embedded Control System




            After a service activity is performed, to reset the   A.7.2   Set Up: Physical Connection
            schedule, do the following:
                                                                                   WARNING:
            •  Select the service activity on the touchscreen
               (Figure A.25).                                        ONLY CONNECT THE CUSTOMER
                                                                     NETWORK        TO    THE     CUSTOMER
            •  Press the Reset button.                               ETHERNET CONNECTION AS SHOWN
                                                                     IN FIGURE A.26.
            •  Enter the password (as explained in Section
               A.4).                                                 DO NOT UNPLUG EXISTING ETHERNET
                                                                     CABLES USED FOR THE INTERNAL
            A.7  Remote Monitoring                                   CONNECTIONS.  DO NOT ATTEMPT
                                                                     TO REMOVE ETHERNET SECURITY
            CAUTION:  The information systems personnel              PROTECTORS TO USE FOR CUSTOMER
            should be notifi ed before changing any of the            ETHERNET CONNECTION.
            network settings.  Changing the settings could
            keep the equipment from working properly.                IF THE CUSTOMER ETHERNET IS
                                                                     PLACED IN AN INTERNAL ETHERNET
            A.7.1   Set Up: Equipment Required                       CONNECTION,         THE       INTERNAL
                                                                     ETHERNET FUNCTIONALITY WILL NOT
            •  PC with an Ethernet connection                        PERFORM AS DESIGNED.

            •  PC with a web browser, such as Microsoft
               Internet Explorer

            •  Cat5 or better Ethernet cable














                                                                       Customer Ethernet Connection
                                                                               on PCB3 Master

                                                                 Figure A.26 Connecting the cable

                                                                 1.  Using a Cat5 Ethernet cable, connect the
                                                                    medical vacuum system to an Ethernet switch
                                                                    or hub.  Connect the cable to the Customer
                                                                    Ethernet Connection (Figure A.26) on PCB3
                                                                    Master board.

                                                                 2.  Verify the green LINK LED on the printed circuit
                                                                    board illuminates.


                                                             A-15                                 4107 9000 96.03