Troubleshooting
This section is intended to serve as a general guide for identifying the potential functional problems which may occur
during the operation of Amico manifolds.
Components removed for maintenance, must be serviced, repaired and tested only by personnel qualified to work on
equipment used in medical service. Only original manufacturer’s parts as supplied by Amico may be used.
ELECTRICAL FAULTS
SYMPTOM CAUSE CORRECTIVE ACTION
No indicator LED’s on front panel a. Power input. 1. Check for 110 to 240 VAC.
illuminated when gas is flowing
b. Fuse Blown. 2. Replace fuse (1 amp).
c. Power present at manifold 3. Replace the circuit board.
circuit board. Check that orange
wire is 12V and blue wire is 5V.
d. Power not present at manifold 4. Replace the power supply.
circuit board.
Both Red LED’s are flashing, but a. All cylinder valves on both 1. Slowly open the cylinder valves on
both banks are full banks are closed. either side and make sure that
both emergency valves on either
side of the manifold are open.