        LINE PRESSURE REGULATOR FAULTS

                     SYMPTOM                             CAUSE                        CORRECTIVE ACTION



         Low pressure relief valve          a. Line regulator set at too high   1. Set line regulator to valve
         venting                                 a delivery pressure.               specified on page 8
                                                                                               (Line Regulators).

                                            b. Relief valve set at too low a   2. Replace the relief valve with a new
                                                 pressure.                         set in accordance with page 9
                                                                                   (Pressure Relief Valves).

                                            c. Creep in line regulator.        3. Replace Line Regulator that
                                                                                    creeps and change seal.








        SHUTTLE VALVE FAULTS


                     SYMPTOM                             CAUSE                         CORRECTIVE ACTION


          Gas leakage                       a. Body halves not joined tightly   1. Tighten screws.
                                                 enough.


                                            b. Body “O” ring leaking.          2. Replace “O” ring kit, use repair kit.

                                            c. Gas bleeding out of shaft of    3. Replace check-valve(s).
                                                shuttle valve, (i.e.: left side
                                                and feeding from the right
                                                high pressure side).

                                            d. Shaft “O” ring leaking, (i.e.: left    4. Use shuttle valve repair kit.
                                                 side, and feeding from the left
                                                 high pressure side).



          Both banks feeding                a. Shuttle valve seat leaking.     1. Replace shuttle valve seal, use
                                                                                    repair kit.

                                            b. Shuttle valve shaft hung up.    2. Bleed gas, reset operating
                                                                                    regulators.






                                                                                                              PAGE   27