PROPRIETARY AND CONFIDENTIAL DRAFT 28/Feb/03
Setup Procedures for Master Alarm
Chapter 1: Introduction
Setup Procedures for Master Alarm
Setting up the Master Alarm involves the following procedures:
• Labeling (see “Master Alarm Labeling” on page 1-54)
• Powering on (see “Master Alarm Power-Up” on page 1-60)
• Configure by using one the following options:
• Front panel keys (see “Configuring the Master Alarm by Using the
Front Panel” on page 1-63)
• Web pages (see “Configuring and Browsing the Master Alarm by
Using the Web Pages” on page 1-72)
Master Alarm Labeling
Tools required: 5/32" hex wrench
Labels (provided)
Input signal data table
During the installation of the MedPlus TotalAlert™ Alarm Network, a data
table should have been completed for the source input signals that are
connected to the Master Alarm (see table 1-27 on page 1-55).
1. If the input signal data table was not completed during the installation,
complete it now by performing the following steps:
a. Record the number of the terminal on the Breakout P.C. Board that
each input is connected to.
b. Record the type of input.
c. Record the LED number that you want to light when that input has an
alarm condition.
NOTE:
Each LED only accepts one type of gas or vacuum.
When you are facing the front panel, the LEDs are numbered one through nine.
1 2 3
4 5 6
7 8 9
Page 1 - 54 MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)