General Operation
1
Chapter 1: Introduction
General Operation
When all systems are operating normally the following occurs:
• The Local Alarm’s green LEDs are lit.
• The Area Alarm Module’s green Normal LED is lit, and the pressure or
vacuum is displayed.
• The Master Alarm lights all the configured LEDs green and scrolls “No
Alarms” across the display’s screen.
Pre-Use Checkout
WARNING:
After the alarm is completely installed, test for cross-connections in
accordance with the latest edition of the NFPA 99 standard or
CSA Z305.1-92 code dictated by the authority having jurisdiction.
Misconnection of medical gas or vacuum can result in serious or fatal
injury to patients.
MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra) Page 1 - 53