Page 693 - Med Plaza and Cancer Center
P. 693

PROPRIETARY AND CONFIDENTIAL DRAFT 28/Feb/03

                   Configuring and Browsing the Master Alarm by Using the Web Pages
                   Chapter 1: Introduction

                          Transfer Configuration

                                 After one Master Alarm is configured, the following information can be
                                 transferred to another Master Alarm across the network:
                                   • Input Information from each configured input
                                      •   Gas-type
                                      •   Message
                                      •   LED configuration
                                      •   System number

                                   • Silenced return time

                                 NOTE:
                                 To comply with the requirements of the City of New York Department of
                                 Buildings; Materials and Equipment Acceptance Division (M.E.A.), do not set
                                 the alarm return time to “never.” Set the alarm to return every 30 minutes.

                                   • SMTP address

                                   •SMTP server
                                   •User names
                                   • User passwords


                                 NOTE:
                                 The electronic notification recipients and the alarm network settings do not
                                 transfer.

                                 1. To transfer the device configuration, click on the Transfer Setup link. The
                                     Configuration Transfer page displays (see figure 1-44 on page 1-97).

































                   Page 1 - 96           MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)