Configuring and Browsing the Master Alarm by Using the Web Pages
1
Chapter 1: Introduction
Figure 1-44. Configuration Transfer
WARNING:
If one Master Alarm is configured to monitor different inputs from the
Master Alarm that you are currently configuring, ensure that the
differences are accounted for in each configuration. Failure to configure
the Master Alarms correctly could endanger patient welfare.
2. Before transferring a configuration, ensure that all of the monitored signals
for both of the Master Alarms should be the same.
NOTE:
The configuration can be transferred to any Master Alarm on this network.
This page lists those that are available.
3. Click on the Master Alarm that you want the configuration to be
transferred to.
NOTE:
The transfer function can only be done across the Ethernet. You cannot transfer
the configuration by using the RS-232 serial port.
MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra) Page 1 - 97