                                                                                        Theory of Operation
                                                                              Chapter 3: Theory of Operation

                              During an alarm condition or a system error on the Master Alarm, the
                              following occurs:

                               • Master Alarm sounds an alarm, and the corresponding red alarm LED
                                 lights, if applicable (errors do not light LEDs).

                               • Message displays.
                               • Alarm or error logs into the Master Alarm’s memory.

                               • Electronic message is sent.
                               • If the Network Communications Board is installed in the Area Alarm, and
                                 the Area Alarm Modules fail to respond to the Master Alarm; the Master
                                 Alarm sounds, message is sent to the event log and displayed, and an
                                 electronic message is sent.
                                                                                                                 3
                              During an alarm condition or a system error on the Area Alarm, the following
                              occurs:
                               • Area Alarm sounds an alarm, and the corresponding red high or low alarm
                                 LED lights, if applicable (errors do not light LEDs).
                               • If the gas pressure is above or below its setpoints or if the vacuum is below
                                 its setpoint, the pressure reading flashes.
                               • If it is an alarm error, the error code displays on the corresponding Area
                                 Alarm Module.

                               • If the Network Communications Board is installed in the Area Alarm:
                                   •  Area Alarm message is sent to the Master Alarm.

                                   •  Master Alarm sounds.
                                   •  Message displays on the Master Alarm.

                                   •  Error or alarm logs into the Master Alarm’s event log.
                                   •  Electronic message is sent from the Master Alarm.

                              All alarms include the following:

                               • Alarm silence key to silence the audible alarm.

                               • Red LEDs that remain lit until the alarm condition is no longer present.
                               • Test key to perform a self-test of each alarm.















               MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)     Page 3 - 5