Page 811 - Med Plaza and Cancer Center
P. 811
PROPRIETARY AND CONFIDENTIAL DRAFT 28/Feb/03
4.3 Master Alarm Battery
Chapter 4: Removal, Replacement, and Adjustment Procedures
4.3 Master Alarm Battery
Tools required: 5/32" hex wrench
Screwdriver
Removal
1. Using a 5/32" hex wrench, unlock the front panel (A) (see figure 4-4 on
page 4-12).
Figure 4-4. Master Alarm Battery
SHOCK HAZARD:
Turn off the power supply in the alarm, and leave it off during the
installation or replacement of any component or sub-assembly. Failure
to do so could result in personal injury or equipment damage.
2. Turn the power switch (D) to the OFF position.
3. Slide a screwdriver underneath the battery (C), and lift it out of its holder
(B).
Page 4 - 12 MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)