Page 811 - Med Plaza and Cancer Center
P. 811

PROPRIETARY AND CONFIDENTIAL DRAFT 28/Feb/03

                   4.3  Master Alarm Battery
                   Chapter 4: Removal, Replacement, and Adjustment Procedures

                   4.3    Master Alarm Battery

                                 Tools required:   5/32" hex wrench
                                                   Screwdriver

                          Removal


                                 1. Using a 5/32" hex wrench, unlock the front panel (A) (see figure 4-4 on
                                     page 4-12).

                                               Figure 4-4. Master Alarm Battery









































                                 SHOCK HAZARD:
                                 Turn off the power supply in the alarm, and leave it off during the
                                 installation or replacement of any component or sub-assembly. Failure
                                 to do so could result in personal injury or equipment damage.

                                 2. Turn the power switch (D) to the OFF position.

                                 3. Slide a screwdriver underneath the battery (C), and lift it out of its holder
                                     (B).







                   Page 4 - 12           MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)