                   4.5  Master Alarm or Area Alarm Transformer
                   Chapter 4: Removal, Replacement, and Adjustment Procedures

                   4.5    Master Alarm or Area Alarm Transformer

                                 Tools required:   5/32" hex wrench
                                                   Phillips head screwdriver
                                                   Needle nose pliers

                          Removal



                                 SHOCK HAZARD:
                                 Turn off the power to the alarm, and leave it off during the installation or
                                 replacement of any component or sub-assembly. Failure to do so could
                                 result in personal injury or equipment damage.

                                 1. Turn off the power to the alarm at the facility’s supply.

                                 2. Using a 5/32" hex wrench, unlock the front panel.


                                 NOTE:
                                 For clarity, the front panel is not shown in the illustration for this procedure.

                                 3. Using a phillips head screwdriver, remove the five screws (A) that secure
                                     the cover plates (B) and (C) to the rough-in assembly (D) (see figure 4-6 on
                                     page 4-17).









































                   Page 4 - 16           MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)