                   4.6  Master Alarm or Area Alarm Conversion Kit Power Supply
                   Chapter 4: Removal, Replacement, and Adjustment Procedures

                          Replacement


                                 1. Install the two screws (H) securing the power supply assembly (C) to the
                                     rough-in box (I).


                                 2. Remove the two nuts (A) securing the high voltage cover (B) to the power
                                     supply assembly (C).

                                 3. Push out the knockout in the high voltage cover (B) where the box
                                     connector (D) goes.


                                 4. Install the box connector (D) in the knockout of the high voltage cover (B).

                                 5. Push the input power wires (G) into the knockout in the high voltage cover
                                     (B), ensuring that there is approximately 1" (25 mm) of sleeving (J) routed
                                     through the box connector (D).




                                 SHOCK HAZARD:
                                 Properly ground the MedPlus TotalAlert™ Alarm Network Conversion
                                 Kit. Failure to do so could cause personal injury or equipment damage.

                                 6. Connect the input power wires (G) to the terminal block (F) in the
                                     following order: line voltage to L, neutral to N, and ground to the earth
                                     ground symbol (K).

                                 7. Tighten the two screws (M) on the box connector (D) to secure the wiring.

                                 8. Tighten the box connector (D) to secure it to the high voltage cover (B).


                                 9. Connect the ground wire (E) to the terminal block (F).

                                 10. Carefully route the ground wire (E) from the terminal block (F) through the
                                     small opening (L) in the high voltage cover (B).




                                 SHOCK HAZARD:
                                 Use care when installing the high voltage cover. Do not pinch or bind
                                 the input power wiring or the ground wire. Personal injury or equipment
                                 damage could occur.

                                 11. Being careful not to pinch or bind any wiring, install the two nuts (A) to
                                     secure the high voltage cover (B) to the power supply assembly (C).

                                 12. Ensure that the power switch is in the ON position.