PROPRIETARY AND CONFIDENTIAL DRAFT 28/Feb/03

                                                                        4.7  Area Alarm Breakout P.C. Board
                                                Chapter 4: Removal, Replacement, and Adjustment Procedures

                              4. Note the positions of the wiring on the connector (I) for the transducers.

                              5. Using a screwdriver, loosen the screws in the transducers wiring connector
                                 (I).


                              6. Remove the transducer wiring.

                              7. Using needle nose pliers, pull the flag terminals (F) off of the Breakout
                                 P.C. Board (C) connectors (G).


                              NOTE:
                              There may be auxiliary wiring to the building automation system that needs to
                              be disconnected.

                              8. Pull the Breakout P.C. Board (C) off of the mounting pegs (D).


                       Replacement


                              1. To replace the Breakout P.C. Board, reverse the order of the removal
                                 instructions.

                              2. To ensure proper operation, perform the “Function Checks” on page 2-5.
                                                                                                                 4











































               MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)    Page 4 - 23