                                                                                    4.9  Area Alarm Module
                                                Chapter 4: Removal, Replacement, and Adjustment Procedures

                                            Figure 4-10. Area Alarm Module



































                                                                                                                 4




                              SHOCK HAZARD:
                              Turn off the power supply in the alarm, and leave it off during the
                              installation or replacement of any component or sub-assembly. Failure
                              to do so could result in personal injury or equipment damage.

                              3. Turn the power switch (G) to the OFF position.

                              4. Push open the ears (E) of the connectors that hold the ribbon cable (B).


                              5. Remove the ribbon cable (B) from the Area Alarm Module (F).

                              6. Using a phillips head screwdriver, remove the two screws (C) from the
                                 retainer ring assembly (D and I).

                              7. Press the retaining clips (H), and pull the retainer ring (I) off of the Area
                                 Alarm Module (F).

                              8. Remove the Area Alarm Module (F) and the retainer ring assembly
                                 (D and I).







               MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)    Page 4 - 27