                   4.10  Blank Area Alarm Module
                   Chapter 4: Removal, Replacement, and Adjustment Procedures

                   4.10 Blank Area Alarm Module

                                 Tools required:   5/32" hex wrench
                                                   Phillips head screwdriver

                          Removal


                                 1. Using a 5/32" hex wrench, unlock the front panel (A) (see figure 4-11 on
                                     page 4-30).

                                            Figure 4-11. Area Alarm Blank Module









































                                 SHOCK HAZARD:
                                 Turn off the power supply in the alarm, and leave it off during the
                                 installation or replacement of any component or sub-assembly. Failure
                                 to do so could result in personal injury or equipment damage.

                                 2. Turn the power switch (D) to the OFF position.

                                 3. Using a phillips head screwdriver, remove the two screws (B) from the
                                     Area Alarm blank module (C).







                   Page 4 - 30           MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)