4.10 Blank Area Alarm Module
Chapter 4: Removal, Replacement, and Adjustment Procedures
4.10 Blank Area Alarm Module
Tools required: 5/32" hex wrench
Phillips head screwdriver
Removal
1. Using a 5/32" hex wrench, unlock the front panel (A) (see figure 4-11 on
page 4-30).
Figure 4-11. Area Alarm Blank Module
SHOCK HAZARD:
Turn off the power supply in the alarm, and leave it off during the
installation or replacement of any component or sub-assembly. Failure
to do so could result in personal injury or equipment damage.
2. Turn the power switch (D) to the OFF position.
3. Using a phillips head screwdriver, remove the two screws (B) from the
Area Alarm blank module (C).
Page 4 - 30 MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)