                   4.12  Local Alarm P.C. Board
                   Chapter 4: Removal, Replacement, and Adjustment Procedures

                   4.12 Local Alarm P.C. Board

                                 Tools required:   Phillips head screwdriver
                                                   5/16" nut driver

                          Removal


                                 1. Using a phillips head screwdriver, loosen the screws (A) enough to allow
                                     the front panel (B) to slide off (see figure 4-13 on page 4-34).

                                             Figure 4-13. Local Alarm P.C. Board







































                                 2. Slide the front panel (B) up and off of the back box (G).



                                 SHOCK HAZARD:
                                 Turn off the power supply in the alarm, and leave it off during the
                                 installation or replacement of any component or sub-assembly. Failure
                                 to do so could result in personal injury or equipment damage.

                                 3. Turn the power switch (H) to the OFF position.


                                 4. Unplug the buzzer wiring (E).