                   4.14  Local Alarm Buzzer Assembly
                   Chapter 4: Removal, Replacement, and Adjustment Procedures

                   4.14 Local Alarm Buzzer Assembly

                                 Tools required:   Phillips head screwdriver


                          Removal

                                 1. Using a phillips head screwdriver, loosen the four screws (A) enough to
                                     allow the front panel (B) to slide off (see figure 4-15 on page 4-38).

                                          Figure 4-15. Local Alarm Buzzer Assembly







































                                 2. Slide the front panel (B) up and off of the back box (G).



                                 SHOCK HAZARD:
                                 Turn off the power supply in the alarm, and leave it off during the
                                 installation of any component or sub-assembly. Failure to do so could
                                 result in personal injury or equipment damage.


                                 3. Turn the power switch (H) to the OFF position.

                                 4. Unplug the buzzer wiring (C).







                   Page 4 - 38           MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)