PROPRIETARY AND CONFIDENTIAL DRAFT 28/Feb/03

                   4.16  Power Supply Fuse Holder
                   Chapter 4: Removal, Replacement, and Adjustment Procedures

                   4.16 Power Supply Fuse Holder

                                 Tools required:   Needle nose pliers
                                                   5/32" hex wrench
                                                   or
                                                   Phillips head screwdriver

                          Removal



                                 SHOCK HAZARD:
                                 Turn off the source power to the alarm, and leave it off during the
                                 installation of any component or sub-assembly. Failure to do so could
                                 result in personal injury or equipment damage.


                                 1. Turn off power to the alarm at the facility’s source.

                                 2. Depending on which alarm you are replacing the fuse on, take one of the
                                     following steps:
                                      •   Unless the replacement is being performed on a Local Alarm, use a
                                          5/32" hex wrench to unlock the front panel (A) (see figure 4-17 on
                                          page 4-43).
                                      •   When changing the power supply fuse holder in the Local Alarm, use
                                          a phillips head screwdriver to remove the front panel.







































                   Page 4 - 42           MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)