                                                                                  4.15  Power Supply Fuse
                                                Chapter 4: Removal, Replacement, and Adjustment Procedures



                              SHOCK HAZARD:
                              Turn off the power supply in the alarm, and leave it off during the
                              installation of any component or sub-assembly. Failure to do so could
                              result in personal injury or equipment damage.

                              2. Turn the power switch (D) to the OFF position.

                              3. Using a screwdriver, remove the fuse cover (B).


                              4. Remove the fuse (C).

                       Replacement



                              CAUTION:
                              Replace the fuse with the same type and rating. If the wrong fuse is
                              installed, equipment may be damaged.

                              1. Ensure the replacement fuse is the same type and rating as the original fuse.

                              2. To replace the fuse, reverse the order of the removal instructions.
                                                                                                                 4
                              3. To ensure proper operation, perform the “Function Checks” on page 2-5.









































               MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)    Page 4 - 41