4.15 Power Supply Fuse
Chapter 4: Removal, Replacement, and Adjustment Procedures
SHOCK HAZARD:
Turn off the power supply in the alarm, and leave it off during the
installation of any component or sub-assembly. Failure to do so could
result in personal injury or equipment damage.
2. Turn the power switch (D) to the OFF position.
3. Using a screwdriver, remove the fuse cover (B).
4. Remove the fuse (C).
Replacement
CAUTION:
Replace the fuse with the same type and rating. If the wrong fuse is
installed, equipment may be damaged.
1. Ensure the replacement fuse is the same type and rating as the original fuse.
2. To replace the fuse, reverse the order of the removal instructions.
4
3. To ensure proper operation, perform the “Function Checks” on page 2-5.