4.11 Local Alarm Front Panel
Chapter 4: Removal, Replacement, and Adjustment Procedures
4.11 Local Alarm Front Panel
Tools required: Phillips head screwdriver
Removal
1. Using a phillips head screwdriver, loosen the four screws (A) enough to
allow the front panel (B) to slide off (see figure 4-12 on page 4-32).
Figure 4-12. Local Alarm Front Panel
2. Slide the front panel (B) up and off of the back box (E).
SHOCK HAZARD:
Turn off the power supply in the alarm, and leave it off during the
installation or replacement of any component or sub-assembly. Failure
to do so could result in personal injury or equipment damage.
3. Turn the power switch (F) to the OFF position.
4. Unplug the buzzer wiring (D).
Page 4 - 32 MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)