Puritan-Bennett Quick-Connect Valve
1. Back off the two mounting screws (1) until the 5. Reassemble the quick-connect valve.
latch valve (2) can be pulled from the outlet. As 6. Reinstall the latch valve in the outlet block. Coat
the screws are loosened, system pressure will the connector (5) with a thin film of silicone
push the latch valve from the outlet. lubricant to aid in insertion. Use care not to
2. Remove the retaining ring (3) from the back of overtighten mounting screws (1) as this can
the valve assembly (2) by removing the two cause distortion of the latch valve.
screws (4). 7. Insert a gas-specific adapter in the latch valve. It
3. Remove the connector (5) from the valve should enter smoothly, latch, flow gas, and
assembly (2). remain in place. If not, replace the entire latch-
4. Remove the spring and poppet (6) and primary valve assembly (2).
seal (7). Obtain a new primary seal (7). Inspect
other items for wear or damage, and replace with
new parts as necessary.
3 4
2 5
6
7
1
Puritan-Bennett Quick-Connect Valve
5
Part No. 847036-00 Rev. D