ZONE VALVE/ AREA ALARM / OUTLET PERFORMANCE/GAS CONCENTRATION AND PURITY INSPECTION DOCUMENT

           Facility   Select Specialty Hospital-Orlando       Inspector  Seth Stephens
          City/State  Orlando, FL                             Date       11-17-17        Page       1 of 2
          Contractor  n/a                                     Facility Contact   Darrell Huckaby

          Area Served   Rooms 8601-8604                       Floor/Bldg   1  Floor
                                                                           st

           Zone Valve Location for this Area   Wall by front entrance   ZVB Manufacturer      Chemetron

         FL Hosp. Tag #          MA/O2/VA-P-1LR-01-B-Z-W-01
          Location IAW NFPA 99   Pass               Labeled      Pass     Gauges     Pass      Leak Test   Pass
           Test Gauge   O2     54   AIR   47    VAC   22
           Pressure vs.
           ZVB Gauge           53         49          22

           Valve Size       ½’’        ½’’         1’’
          Comments: *NOTE: ZVB/ALARM NEEDS FH STYLE TAGS ON THE OUTSIDE OF THE WINDOW.


         Gas/Pipeline Purity  Oxygen      Med Air       Comments
         Hydrocarbons             <1 PPM         <1 PPM   NOTE: RAISE PRESSURE FOR MEDICAL AIR (AT 47 NOW SHOULD
         Dew Point                    -12            23                                            BE 50 OR ABOVE).
         Carbon Monoxide          <1 PPM         <1 PPM
         Carbon Dioxide           <1 PPM        410 PPM
         Gas Concentration        99%O2       21.5(%O2)
         Particulates                 <1             <1

         Odor/Discoloration         None           None
         Test Location              8603                  8603

         Area Alarm Panel Location for this Area         Nurses station

         Location IAW NFPA 99       Pass                 Labeled    Pass   Gauges     Pass     Leak Test   Pass
         Area Alarm Panel Manufacture Data         Amico Alert (double) GP-L-1LR-01-6G-003
                     GAUGE     LOW      HIGH                        TEST
          GAS/VAC      PSI    ALARM    ALARM    VISUAL   AUDIBLE   SWITCH   SENSOR        COMMENTS        PASS/FAIL
            OXY        54       40       60       P        P         P         PS                           PASS
             AIR       48       40       60       P        P         P         PS                           PASS
            VAC        22       12       N/A      P        P         P         PS                           PASS
           PS= Patient Side   SS= Source Side   NW= Not Working   CNT= Could Not Test   NA = Not Applicable    NP = Not Present