MEDICAL AIR SYSTEM INSPECTION FORM

          Facility            Select Specialty Hospital-Orlando   Technician      Anthony Voight
          City/State          Orlando, FL                      Date               12-12-17

          System Inspected    Med Air System                   Floor/Bldg         Basement
          Manufacturer        Beacon Medaes                    Equipment ID #     CM-H-ICR-GR-0-F-1A1/1B1
          Model#              LPS-7D-SD80-DAY-003              Serial #           65448

          Compressor Configuration    Duplex Stack scroll      Motor HP           7.5HP
          Hours Comp# 1     16,546      Comp# 2     16,331     Comp# 3     -           Comp# 4     -

                  Is Medical Air System located IAW NFPA 99 (Dedicated Mechanical Equipment Room, Adequately Ventilated, with Required
           PASS   Utilities, Minimum 1-HR Fire Rated)
           PASS   Is Medical Air System Fed from Emergency Power (Essential Electrical System)
                  Do the Medical Air Compressors meet the requirements of NFPA 99 (Elimination of oil anywhere in the compressor,
           PASS   separation of the oil containing section by seal(s) with atmospheric vents
                  Is the Medical Air System connected ONLY to the Medical Air Distribution System and used ONLY in the application of Human
           PASS   Respiration, and calibration of Medical Devices for Human Respiration

           PASS   Does the Medical Air System consist of at least (2) Two Compressors
           PASS   Is the Medical Air Intake & Discharge connected to the Piping by Flexible Connectors
           PASS   Is the Medical Air Intake located IAW NFPA 99
           PASS   Is there an Discharge Isolation Valve, Check Valve and Pressure Relief Valve for each Compressor

           PASS   Is there at least Duplex After Coolers w/Condensate Drain Traps for the System
                  Is the Medical Air Receiver Corrosion Resistant & equipped with a Pressure Relief Valve, Automatic Drain, Manual Drain, Sight
           PASS   Glass, Pressure Indicator and Valved Bypass Arrangement

           PASS   Is the Medical Air System equipped with at least Duplex Dryers    Desiccant                 x   Refrigerant
           PASS   Is the Medical Air System Equipped with at least Duplex Final Line Regulators with Pressure Indicators
           PASS   Is the Medical Air System Equipped with at Least Duplex Final Line Filters with Continuous Visual Indicator
           PASS   Can the Medical Air Compressors, Dryers, Filters & Regulators be Isolated for Service without shutting down the system

           PASS   Is the Medical Air System Equipped with a ¼” Sample Port Valve and Source Shut Off Valve
           PASS   Is the Medical Air System Equipped with a Dew Point & Co Monitor

           PASS   Are the Local Alarms Required by NFPA 99 Present, Functioning and connected to the Master Alarms as required
          Comments: