MEDICAL VACUUM SYSTEM ANNUAL INSPECTION FORM

          Facility           Select Specialty Hospital-Orlando     Technician  Anthony Voight
          City/State         Orlando, FL                           Date        12-12-17

          System Inspected  Med VAC                                Floor/Bldg   Basement
          Manufacturer       Beacon/Busch                          Equipment ID #
          Model#             LVS-5D-HA120-D4                       Serial #              146474-1

          Vacuum System Configuration     Duplex stack vertical tank  Motor HP           5 HP
          Hours Pump#1     5,651    Pump#2    5,656      Pump#3  --              Pump#4        --

                  Is Medical Vacuum System located IAW NFPA 99 (Dedicated Mechanical Equipment Room, Adequately Ventilated, with
           PASS   Required Utilities, Minimum 1-HR Fire Rated)
           PASS   Is Medical Vacuum System Fed from Emergency Power (Essential Electrical System)

           PASS   Does the Medical Vacuum System consist of at least (2) Two Pumps
           PASS   Is the Medical Vacuum Intake & Discharge connected to the Piping by Flexible Connectors
            NO    Is the Medical Vacuum System used for Analysis, Research or Teaching Laboratory
                  Does the Medical Vacuum System that is used for Analysis, Research or Teaching Laboratory functions have the appropriate
            N/A   Valves and Fluid Trap IAW NFPA 99
            NO    Is the Medical Vacuum System is used for Waste Anesthetic Gas Disposal (WAGD) functions
                  Does the Medical Vacuum System that is used for Waste Anesthetic Gas Disposal (WAGD) functions meet the requirements of
            N/A   NFPA 99 (5.1.3.7.2.1)
           PASS   Is the Medical Vacuum Exhaust located IAW NFPA 99
           PASS   Is there an Inlet Isolation Valve & Check Valve for each Pump

           PASS   Is the Medical Vacuum Receiver equipped with a Manual Drain and Valved Bypass Arrangement
           PASS   Is the Medical Vacuum System Equipped with a Source Shut Off Valve
           PASS   Are the Local Alarms Required by NFPA 99 Present, Functioning and connected to the Master Alarms as required
          Comments: