Page 39 - Select Specialty
P. 39

"Oil-Less” Scroll Medical Air                                                       Medical Systems


               6.0 Inspection/Replacement Procedures



               6.1 V-Belts

               Narrow type V-belts are used for this unit.  Refer to Table 6-1a & 6-1b for the correct size.

               6.1.1 Tension Check

               Table 6-1a Belt size and Tension (for new belts)
                 Hp   Motor Sheave Diameter      Belt Size        TENSIONING        No. of
                      (see note 1)   (see note 2)   (see note 1)   (see note 2) Defl. (in.) Force (lbs.) Belts
                 2       4.17      3.15     SPZ925    3V X 375    0.25       5        1
                 3       2.95      4.50     SPZ875    3V X 400    0.25       6        1
                       Notes:
                         1.  Supplied on units with serial number 262296 and greater.
                         2.  Units with serial numbers prior to 262296.

               Table 6-1b Belt size and Tension (for new belts)
                 Hp   Motor Sheave Diameter      Belt Size        TENSIONING        No. of
                      (see note 1)   (see note 2)   (see note 1)   (see note 2) Defl. (in.) Force (lbs.) Belts
                 5       3.35      3.65     SPZ 987   3V X 450    0.25       5        2
                7 ½           5.00               3V X 475         0.25       6        2
                 10      3.35      3.65     SPZ 1060  3V X 450    0.25       5        4
                 15           5.00               3V X 475         0.25       6        4
                       Notes:
                         1.   Supplied on units with serial number 188378 and greater.
                         2.   Units with serial numbers prior to 188378.


                                                           WARNING:
                        BEFORE STARTING ANY MAINTENANCE PROCEDURES, DISCONNECT ALL POWER
                                                       TO THE PACKAGE.


               Release all pressure from the package before removing, loosening, or servicing any covers, guards, fittings,
               connections, or other devices.

               Never perform any maintenance functions while the unit is in operation.

               Check the belt tension monthly.  Disconnect the main power and remove the beltguard. As
               shown in the illustration at right, deflect each V-belt at the center of the drive span with a
               spring balance or tension meter at the tension force of Table 6-1. Then check that the average
               deflections at the proper tension force are approximately the same values as shown in Table 6-
               1.