"Oil-Less” Scroll Medical Air Medical Systems
5.0 Maintenance
5.1 Maintenance Schedule
WARNING:
BEFORE STARTING ANY MAINTENANCE PROCEDURES, DISCONNECT ALL POWER
TO THE PACKAGE.
Release all pressure from the package before removing, loosening, or servicing any covers, guards,
fittings, connections, or other devices.
Never perform any maintenance functions while the unit is in operation.
Maintenance Schedule
Item Frequency Action
Check condensate in tank Daily Open manual drain valve: check auto
drain
Check operation of safety valve Weekly Manually release pressure
Check inlet air filter(s) Weekly Inspect and clean or replace
Check nuts, bolts, fittings, etc. Monthly Inspect and tighten
Check belt tension Monthly Inspect and tighten or replace
Check flow through orifice of dew Every 6 months Check for flow blockage
point sensor
Check dew point sensor accuracy Every 12 months Verify dew point sensor accuracy
(contact BeaconMedæs)
Calibrate CO monitor Every 6 months See Appendix B
Replace pre-filters & afterfilters Yearly Replace filter elements
Replace Dryer Desiccant 5 years See Appendix A
Orbiting & Drive Bearing Lubrication * 10,000 Hours Contact BeaconMedæs Customer
(2, 3, 5 & 10HP Scrolls Only) Service Department
Replace Compressor Tip Seals * 10,000 Hours Contact BeaconMedæs Customer
Service Department
*An authorized BeaconMedæs Service Technician must perform compressor tip seals replacement.
This replacement will be done by BeaconMedæs at no charge to the customer. Please contact
BeaconMedæs Customer Service department at 1-800-756-2590 prior to 10,000 operating hours to
schedule this maintenance. Please have the serial number of your compressor system available.
5-1