Page 37 - Select Specialty
P. 37

"Oil-Less” Scroll Medical Air                                                       Medical Systems


               5.0 Maintenance



               5.1 Maintenance Schedule

                                                           WARNING:
                    BEFORE STARTING ANY MAINTENANCE PROCEDURES, DISCONNECT ALL POWER
                                                      TO THE PACKAGE.


               Release all pressure from the package before removing, loosening, or servicing any covers, guards,
               fittings, connections, or other devices.

               Never perform any maintenance functions while the unit is in operation.

               Maintenance Schedule
                Item                                Frequency              Action
                Check condensate in tank            Daily                  Open manual drain valve: check auto
                                                                           drain
                Check operation of safety valve     Weekly                 Manually release pressure

                Check inlet air filter(s)           Weekly                 Inspect and clean or replace

                Check nuts, bolts, fittings, etc.   Monthly                Inspect and tighten

                Check belt tension                  Monthly                Inspect and tighten or replace

                Check flow through orifice of dew   Every 6 months         Check for flow blockage
                point sensor

                Check dew point sensor accuracy     Every 12 months        Verify dew point sensor accuracy
                                                                           (contact BeaconMedæs)
                Calibrate CO monitor                Every 6 months         See Appendix B

                Replace pre-filters & afterfilters   Yearly                Replace filter elements

                Replace Dryer Desiccant             5 years                See Appendix A

                Orbiting & Drive Bearing Lubrication *  10,000 Hours       Contact BeaconMedæs Customer
                (2, 3, 5 & 10HP Scrolls Only)                              Service Department
                Replace Compressor Tip Seals *      10,000 Hours           Contact BeaconMedæs Customer
                                                                           Service Department

               *An authorized BeaconMedæs Service Technician must perform compressor tip seals replacement.
               This replacement will be done by BeaconMedæs at no charge to the customer.  Please contact
               BeaconMedæs Customer Service department at 1-800-756-2590 prior to 10,000 operating hours to
               schedule this maintenance. Please have the serial number of your compressor system available.





                                                             5-1
   32   33   34   35   36   37   38   39   40   41   42