Page 11 - Womens Pavilion
P. 11

MEDICAL VACUUM SYSTEM ANNUAL INSPECTION FORM

          Facility            Florida Hospital Orlando-Women’s     Technician     Anthony Voight
          City/State          Orlando, FL                          Date           12-11-17

          System Inspected    Med VAC #7                           Floor/Bldg     Ground Floor Women’s
          Manufacturer        Beacon                               Equipment ID #    No FL Hosp. #
          Model#              4107108709                           Serial #          HOP300755

          Vacuum System Configuration     Triplex Claw             Motor HP          15 HP
          Hours Pump#1  5,185        Pump#2     5,193      Pump#3    1,103     Pump#4         -

                  Is Medical Vacuum System located IAW NFPA 99 (Dedicated Mechanical Equipment Room, Adequately Ventilated, with
           YES    Required Utilities, Minimum 1-HR Fire Rated)
           YES    Is Medical Vacuum System Fed from Emergency Power (Essential Electrical System)
           YES    Does the Medical Vacuum System consist of at least (2) Two Pumps
           YES    Is the Medical Vacuum Intake & Discharge connected to the Piping by Flexible Connectors
            n/a   Is the Medical Vacuum System used for Analysis, Research or Teaching Laboratory
                  Does the Medical Vacuum System that is used for Analysis, Research or Teaching Laboratory functions have the appropriate
            n/a   Valves and Fluid Trap IAW NFPA 99
           YES    Is the Medical Vacuum System is used for Waste Anesthetic Gas Disposal (WAGD) functions

                  Does the Medical Vacuum System that is used for Waste Anesthetic Gas Disposal (WAGD) functions meet the requirements of
           YES    NFPA 99 (5.1.3.7.2.1)
           YES    Is the Medical Vacuum Exhaust located IAW NFPA 99
           YES    Is there an Inlet Isolation Valve & Check Valve for each Pump
           YES    Is the Medical Vacuum Receiver equipped with a Manual Drain and Valved Bypass Arrangement
           YES    Is the Medical Vacuum System Equipped with a Source Shut Off Valve
           YES    Are the Local Alarms Required by NFPA 99 Present, Functioning and connected to the Master Alarms as required

          Comments:












              Corporate Office ∙ 4518 Clements Rd. Lakeland Florida 33811 Phone: 863-646-7555 ∙ Fax 863-646-2999
                                 “Medical Gas Pipeline Equipment Service, Sales & Verifications”
                                                      ©Daniel Voight 2009
   6   7   8   9   10   11   12   13   14   15   16