Page 12 - Womens Pavilion
P. 12

ZONE VALVE/ AREA ALARM / OUTLET PERFORMANCE/GAS CONCENTRATION AND PURITY INSPECTION DOCUMENT

           Facility   Florida Hospital Orlando                Inspector  Steve Gourley
          City/State  Orlando, Fl.                            Date       12-5-17         Page       1 of 2
          Contractor  n/a                                     Facility Contact   Darrell Huckaby

          Area Served   Bio Med                               Floor/Bldg   Ground Floor Womens

           Zone Valve Location for this Area   Outside of Bio Med     ZVB Manufacturer        Beacon

           FL Hosp. Tag #        O2 #000000222352  Air #000000222353
          Location IAW NFPA 99   Pass               Labeled      Pass     Gauges     Pass      Leak Test   Pass
           Test Gauge   O2     52   AIR   49
           Pressure vs.
           ZVB Gauge           50         52

           Valve Size       ¾”         ¾”
          Comments:


         Gas/Pipeline Purity  Oxygen      Med Air       Comments
         Hydrocarbons             <1 PPM        <1 PPM
         Dew Point                   -44F            5F
         Carbon Monoxide          <1 PPM        <1 PPM
         Carbon Dioxide            0 PPM       120 PPM
         Gas Concentration        99%O2       21.8(%O2)
         Particulates                 <1             <1

         Odor/Discoloration         none           none
         Test Location            Bio Med       Bio Med

         Area Alarm Panel Location for this Area         No alarm at this location












              Corporate Office ∙ 4518 Clements Rd. Lakeland Florida 33811 Phone: 863-646-7555 ∙ Fax 863-646-2999
                                 “Medical Gas Pipeline Equipment Service, Sales & Verifications”
                                                      ©Daniel Voight 2009
   7   8   9   10   11   12   13   14   15   16   17