ZONE VALVE/ AREA ALARM / OUTLET PERFORMANCE/GAS CONCENTRATION AND PURITY INSPECTION DOCUMENT

           Facility    Florida Hosp. Orlando – Women’s Center  Inspector  Anthony Voight
           City/State   Orlando, FL                           Date       11-29-17        Page       1 of 2

          Contactor    n/a                                    Facility Contact   Darrell Huckaby
                                                                           rd
          Area Served   Anesthesia Work Room                  Floor/Bldg   3  Floor

           Zone Valve Location for this Area   Corridor wall, outside of OR area   ZVB Manufacturer   Beacon

          FL Hosp. Tag #       None
          Location IAW NFPA 99   Pass          Labeled    Pass      Gauges   Pass       Leak Test   Pass
           Test Gauge   O2    52   AIR    48   VAC    21
           Pressure vs.
           ZVB Gauge          50          50          21

          Valve Size       ½”          ½”          1”
          Comments:


         Gas/Pipeline Purity  Oxygen        Med Air       Comments
         Hydrocarbons               <1 PPM        <1 PPM
         Dew Point                     -44F            5F
         Carbon Monoxide            <1 PPM        <1 PPM
         Carbon Dioxide              0 PPM       120 PPM
         Gas Concentration          99%O2       21.8(%O2)
         Particulates                 None          None
         Odor/Discoloration           None          None

         Test Location        Anest. Wk. Rm  Anest. Wk. Rm

                                                     No Area Alarm Panel















              Corporate Office ∙ 4518 Clements Rd. Lakeland Florida 33811 Phone: 863-646-7555 ∙ Fax 863-646-2999
                                 “Medical Gas Pipeline Equipment Service, Sales & Verifications”