ZONE VALVE/ AREA ALARM / OUTLET PERFORMANCE/GAS CONCENTRATION AND PURITY INSPECTION DOCUMENT

           Facility    Florida Hosp. Orlando – Women’s Center  Inspector  Anthony Voight
           City/State   Orlando, FL                           Date       11-29-17        Page       1 of 2

           Contractor  n/a                                    Facility Contact   Darrell Huckaby
                                                                           rd
          Area Served   C-Section 1                           Floor/Bldg   3  Floor

           Zone Valve Location for this Area   Corridor wall outside of C-Section 1   ZVB Manufacturer   Beacon

          FL Hosp. Tag #       O2 000000222009 AIR 000000222012 VAC 000000222011 N2O 000000222013
          Location IAW NFPA 99   Pass          Labeled    Pass      Gauges   Pass       Leak Test   Pass
           Test Gauge   O2    52   AIR    48   VAC    21   N2O    54
           Pressure vs.
           ZVB Gauge          54          48          21          55

          Valve Size       ¾”          ½”         1 ¼”        ½”
          Comments:


         Gas/Pipeline Purity  Oxygen       Med Air      N2O           Comments
         Hydrocarbons              <1 PPM        <1 PPM       <1 PPM
         Dew Point                    -44F           5F          -40F
         Carbon Monoxide           <1 PPM        <1 PPM       <1 PPM
         Carbon Dioxide             0 PPM       120 PPM        0 PPM
         Gas Concentration         99%O2      21.8(%O2)     99.9%N2O
         Particulates                None          None         None

         Odor/Discoloration          None          None         None
         Test Location            C-Sect 1      C-Sect 1      C-Sect 1

         Area Alarm Panel Location for this Area         Nurses station across from C-Section 1

         Location IAW NFPA 99       Pass                 Labeled    Pass   Gauges     Pass     Leak Test   Pass
         Area Alarm Panel Manufacture Data         Beacon Total Alert 2 ADV# 000000222142
                     GAUGE     LOW      HIGH                        TEST
          GAS/VAC      PSI    ALARM    ALARM    VISUAL   AUDIBLE   SWITCH   SENSOR        COMMENTS        PASS/FAIL
            OXY        52       40       60       P        P         P         SS                           PASS
             AIR       48       40       60       P        P         P         SS                           PASS
            VAC        21       12       NA       P        P         P         SS                           PASS
            N2O        52       40       60       P        P         P         SS                           PASS
           PS= Patient Side   SS= Source Side   NW= Not Working   CNT= Could Not Test   NA = Not Applicable    NP = Not Present




              Corporate Office ∙ 4518 Clements Rd. Lakeland Florida 33811 Phone: 863-646-7555 ∙ Fax 863-646-2999
                                 “Medical Gas Pipeline Equipment Service, Sales & Verifications”