TotalAlert Infinity  NFPA Medical Gas Notification System
                                   ™


            7.0 Troubleshooting



             Symptom                  Possible Cause                      Corrective Action
             No LED power indicator   AC power is not turned on           Check AC power source.
             illuminated on alarm     Blown power supply fuse             Replace fuse.
             front panel
                                      AC power wiring is not connected    Check AC entrance wiring at power supply
                                                                          terminals.
                                      Power supply DC wire harness is not   Check connections on each end including
                                      connected properly between the      orientation of cable connectors. Reconnect
                                      power supply and the main board     if necessary.
                                      Faulty power supply assembly        Replace power supply
             No green “heartbeat”     Board has not been initialized      Initialize board per instructions in manual.
             power indicator is il-   Board dip switches are not set cor-  Set dip switches per diagram on board
             luminated on connected   rectly.                             label.
             boards
                                      Cable not connected correctly       Check cable connections between boards.
             No green “heartbeat”     Sensor not connected to the Gas     Verify correct sensor connection to Gas
             power indicator is illumi-  Input Board.                     Input Board.
             nated on gas sensor      Gas Input Board does not have power  Verify that Gas Input Board has been initial-
                                      (no “heartbeat”)                    ized and that dip switches are set correctly.
                                                                          See item 2 above.
             Fast green "heartbeat"   Sensor is reading out of range      Verify gas pressure in pipeline. If normal,
             power indicator is illu-                                     replace sensor.
             minated on gas sensor (2
             beats per second)
             Red LED illuminated      Active warning                      Navigate to the warning screen by press-
             above warning indicator                                      ing the Information button, then selecting
             ( ! ).                                                       Warnings.
             Lost Connection          Alarm has lost network communica-   Correct communication problem with
             Warning                  tion with manually enrolled device  networked device or remove the enrolled
                                                                          device from the Enrolled Devices list.
                                      The alarm has lost connection to a   Check connections and cable to board.
                                      connected/ initialized board.
             Input active and         Master signal input is wired, but not   Assign the wired input to a valid alarm
             Unassigned Warning       assigned/programmed to an alarm     signal.
                                      signal
             Sensor active and        Sensor is wired to the sensor input   Assign the wired gas sensor to a valid gas
             Unassigned Warning       board, but not assigned/programmed input.
             F1 displayed on Area     Gas sensor is not connected or has   Check gas sensor connection. If connection
             badge                    shorted.                            is good, replace sensor.
             F2 displayed on Area     Gas sensor is reading below range   Verify gas sensor connections and pipeline
             badge                    (fault low).                        pressure. If connection and pressure are
                                                                          good, replace sensor.
             F3 displayed on Area     Gas sensor is reading above range   Verify gas sensor connections and pipeline
             Badge                    (fault high).                       pressure. If connection and pressure are
                                                                          good, replace sensor.
             F4 displayed on Area     Gas sensor connected is wrong type   Connect the correct type gas sensor.
             Badge                    of gas.