Care of Flexible Endoscopes


                   Not all endoscopes are validated for all the various types of automatic reprocessing
                   machines (AER, ECR, STERIS System 1E, etc.). Always refer to the endoscope
                   manufacturer’s instructions to ensure that the machine is safe to use for the flexible
                   endoscope.

                   10. Drying the Channels Prior to Assembly

                   Dry the channels with a filtered, compressed air device or a syringe. Be sure to stay
                   within the pressure limits set forth by the manufacturer for a compressed air device.
                   Pressure that is too high can tear the interior wall of the channel and cause damage later
                   on. It is recommended to flush the channels with denatured alcohol for optimal drying.
                   Dry the outside of the endoscope with a soft cloth.

                   A drying cabinet also can be used to dry the endoscope (see Figure 7). Hang the flexible
                   endoscope in a vertical position and leave it in the cabinet for the time recommended by
                   the cabinet’s manufacturer.



                   Figure 7 – Hang a Flexible Endoscope in a Drying Cabinet