Page 7 - The Care and Handling of Flexible Scopes v3
P. 7

Care of Flexible Endoscopes


                   Never use sterile saline or saline solution of any kind to clean an endoscope! Saline can
                   damage the metal and epoxy that hold the lenses in place.

                   4.  Transport After Procedure to a Terminal Reprocessing Area

                   Flexible endoscopes can be used in various facility patient care areas that are
                   unaccustomed to using and transporting flexible endoscopes. These patient care areas
                   may be using the endoscope during off-hours or weekends (eg, Emergency Department,
                   Respiratory Department, or CCU to list a few). All of these variables present a challenge
                   to ensuring the flexible endoscope is used as intended, pre-cleaned after the procedure
                   and returned in a timely fashion to reprocess. If the flexible endoscope is left
                   unattended for any length of time, it is subject to possible damage and the debris from
                   the procedure may dry and cake onto the inside of the channels.

                   Some facilities may recommend leaving the endoscope to soak in an enzymatic
                   detergent solution until it can be transferred back to CSSD for reprocessing. Before
                   implementing this step, ensure that the manufacturer has approved prolonged soaking
                   in any solution.

                   The best process is to pre-clean the endoscope in the procedure area, and if transport is
                   delayed, place the endoscope in a transfer bag to keep it moist. The transfer bag should
                   be secured in a basket or tray, separated from other articles for transport. Arrange for
                   transfer as soon as possible to avoid any possible damage and debris drying onto the
                   fiberscope’s surface or channels.



                   5.  Leakage Testing

                   Leakage testing is the very first step that must be done before proceeding with any
                   cleaning steps (see Figure 3). The leakage test will demonstrate if there is a hole in the
                   covering or inner lining of the flexible endoscope. This test needs to be performed after
                   each use AND before the flexible endoscope is cleaned or submersed. If there is a leak in
                   the flexible covering, fluids can seep into the channels that house the viewing optics, the
                   light fibre optics, and the cables that control the distal tip. If that happens, the
                   endoscope can be further damaged and may require costlier repairs.
                   Each flexible endoscope manufacturer recommends using the manual leakage tester
                   designed for its particular flexible fiberscope. Ensure that you are using the correct
                   equipment and process steps recommended by the manufacturer; there is no universal
                   equipment to leak-test a flexible endoscope!
















               Avensys UK Ltd                                                                     Page | 7
   2   3   4   5   6   7   8   9   10   11   12