                                                                                      10
                                                                                                 Garanzia
                                                                                                   N°       Problema           Possibile causa          Soluzione
                                                                                                                           Cortocircuito dovuto a un eccesso di   Asciugare la cavità di accesso con un
              Guidance and Manufacturer's Declaration - Electromagnetic Immunity                                           fluido nella camera pulpare   batuffolo di cotone / getto di aria. In
                                                                                                                           (soluzione di risciacquo, saliva,   caso di emorragia abbondante,
                   ®
        The ProPex  II is intended for use in the electromagnetic environment specified                La reazione del display è   sangue);       attendere che questa si sia arrestata.
                                                                                                                                                  Per l'isolamento:
                                              ®
        below. The customer or the user of the ProPex  II should ensure that it is used in             ipersensibile: L’indicazione   Un contatto diretto della lima con la   - otturazione con preparazione
                                                                                                                           gengiva o proliferazioni gengivali, ad
        such an environment.                                                                     8.    “apex”/”over” (apice/oltre) viene   esempio una corona di metallo   adeguata;
                                                                                                                                                  - utilizzare sempre una diga di
                                                                                                       attivata prima del reale
                        IEC     Compliance                                                             raggiungimento.     fratturato.            gomma.
        Immunity test  60601-1-2   level    Electromagnetic environment - guidance                                         Un contatto diretto della lima con   Isolare la lima posizionandovi sopra
                                                                                                                                                  2-3 stopper in silicone oppure prima
                                                                                                                           otturazioni in metallo (corona, perno
                     Test level                                                                                            parapulpare, riempimento di   dell'uso inserire la lima in un piccolo
                                                                                                                           amalgama).             tubo in polivinile.
                                            Portable and mobile RF communications
                                            equipment should be used no closer to
                                                               ®
                                            any part of the ProPex  II, including
                                            cables, than the recommended
                                            separation distance calculated from the   11         11      GARANZIA
                                            equation applicable to the frequency of                              ®
        Conducted    3 Vrms     3 Vrms      the transmitter.                                             ProPex  II ha una garanzia di 24 mesi a partire dalla data dell’acquisto. Gli
                                                                                                         accessori (cavi, batteria, ecc.) hanno una garanzia di 6 mesi a partire dalla
        disturbances   150 kHz to   150 kHz to                                                           data dell’acquisto.
        induced by   80 MHz     80 MHz      Recommended separation distance:
        RF fields                           d = 1.17 √P                                                  La garanzia è valida per le normali condizioni d'uso. Qualsiasi modifica o
                                            d = 1.17 √P    80 MHz to 800 MHz                             danno accidentale renderà nulla la garanzia.
                                            d = 2.3 √P    800 MHz to 2.7 GHz
                     6 Vrms in   6 Vrms in
                     ISM bands   ISM bands   Where P is the maximum output power
        IEC         150 kHz to   150 kHz to   rating of the transmitter in watts (W)            12      SMALTIMENTO DEL PRODOTTO
        61000-4-6    80 MHz     80 MHz      according to the transmitter
                     80% AM at   80% AM at   manufacturer and d is the recommended                             ATTENZIONE NON GETTARE !!
                     1 kHz      1 kHz       separation distance in meters (m).                                 Questo prodotto e tutti i suoi componenti devono essere
                                                                                                               assolutamente riciclati dal vostro fornitore.
                                            Field strengths from fixed RF
                                            transmitters, as determined by an
                                                                  a
        Radiated RF   10 V/m    10 V/m      electromagnetic site survey , should be
        IEC         80 MHz to              less than the compliance level in each
                                                         b
        61000-4-3    2.7 GHz                frequency range .
                                            Interference may occur in the vicinity of
                                            equipment marked with the following
                                            symbol:









       170/172                           B EU PIIL DFU PRT / Rev.07 / 02-2018 (Old ZF 190292.X)  B EU PIIL DFU PRT / Rev.07 / 02-2018 (Old ZF 190292.X)            155/172