Page 123 - The Veterinary Laboratory and Field Manual 3rd Edition
P. 123

92  Willy Schauwers

            Table 2.4  Plastic ware properties: chemical resistance and possibility to autoclave.

            Plastic ware                      Chemical resistance   Autoclavable
            Polypropylene (PP)                Good                  Yes
            Low-density polyethylene (LDPE)   Good                  No
            High-density polyethylene (HDPE)  Good                  No
            Polymethylpentene (PMP, TPX)      Good                  Yes
            Polycarbonate (PC)                Moderate              Yes
            Polytetrafluoroethylene (PTFE)    Very good             Yes
            Polymethylmethacrylate (PMMA)     Moderate              No
            Polystyrene (PS)                  Moderate              No
            Polyvinylchloride (PVC)           Moderate              No


            volumetric accuracy                      The following information may also be added:


            Volumetric glassware is manufactured and cali-  •  country of origin
            brated in accordance with international ISO   •  error limit
            standards to permit very accurate determina-  •  standard: the standard to which the product
            tion and measurement of specific volumes. All   conforms
            of the volumetric glassware is marked with a set   •  certified glassware bears a serial number/
            of inscriptions in accordance with any specific   batch number for identification and trace-
            standard associated with it.               ability.
              Mandatory inscriptions include:
                                                     General guidelines for use include the following.
            •  manufacturer
            •  class:                                •  Ensure that all volumetric glassware is kept

                                                       scrupulously clean. Dirt, and especially
              •  class A/AS – highest level of accuracy (S
                 stands for swift delivery of pipettes and   grease, can distort the shape of the meniscus
                 burettes), glass or plastic           and also cause droplets of liquid to adhere to
              •  class B – general purpose work calibrated   the vessel walls. Both seriously impair accu-
                 to a lower level of accuracy, glass or plastic  racy (good cleanliness is indicated by uniform
                                                       wetting of the glass surface with distilled
            •  nominal volume                          water).
            •  volume unit                           •  Never pipette by mouth. Always use a pur-
            •  calibration temperature                 pose designed pipette filler.
            •  for pipettes:                         •  Autoclaving at 121°C and cleaning in auto-
                                                       matic dishwashers is acceptable and will not
              •  IN: calibrated to contain, or         affect the accuracy of glassware.
              •  EX: calibrated to deliver           •  All items should be held in a vertical position
              •  waiting time                          when reading the meniscus. The menis-
              •  blowout: indication that last drop should   cus should be at eye level to avoid parallax
                 be blown out of jet.                  errors.